Using a blank formulation and a known spike
Prepare the same beverage base with and without added creatine. Test recovery by adding a known amount of creatine standard. Keep dilution or extraction steps consistent and check that sample preparation neither converts creatine nor causes it to be lost.
Distinguishing creatine from creatinine
Check whether the spike increases the signal by the expected amount. Even good recovery cannot demonstrate selectivity if creatinine contributes to the same signal. Dash and Sawhney describe a liquid chromatography method that measures the two compounds separately. Verify its performance in the beverage being tested.